Regulatory Information
General Specifications
Environmental
Operating Temperature:
Storage Temperature:
Humidity:
Altitude:
EMC:
Physical
Weight:
Dimensions: (height x width
x depth)
Power
Operating Voltage Range:
Requirements
Operating Frequency
Range:
Power Dissipation:
Cooling
To provide adequate cooling, and air gap of approximately 75mm (3ins) should
Requirements
be maintained around the vented sections of the instrument.
Cleaning
Use a soft, clean, damp cloth to clean the front-panel and side covers.
Use
This instrument is designed for indoor use only.
WARNING
Appliance coupler (mains input power cord) is the power disconnect device. Do
not position the instrument such that access to the coupler is impaired.
WARNING
For continued protection against fire hazard, replace the line fuse only with the
same type and line rating (250V, F3.15A, 20mm fast blow fuse with high
breaking capacity, Keysight part number 2110-0957).
The use of other fuses or materials is prohibited.
36
General Specifications
0 to +55°C
−20 to +70°C
Up to 95% Relative Humidity to +40°C
3000m (9,840 ft.)
M
EN55011: 1991 (G
EETS
C
A),
EN50082-1:1992
LASS
AND
4.2 kg nominal
130H x 250W x 462D mm nominal
(including handle and rubber bumpers)
88H x 212W x 363D mm nominal
(excluding handle and rubber bumpers)
85 - 264 Vac
47 - 440Hz
50 VA (maximum)
EPM-P Installation Guide