Download Print this page

General Specifications - Keysight U1583B Operating Instructions Manual

Current clamp

Advertisement

-
-
-
-
-
-
-
-
-
-
-
-
-
-
Standard Items Purchase Checklist
The following items are included when you make a purchase:
-
U1583B current clamp
-
BNC-to-banana plug adapter
-
Operating Instructions (this sheet)

General Specifications

The current clamp is designed for indoor use.
Specification
Specified Current Range
Usable Current Range
AC crest factor
Bandwidth
Weight
Cable length
Maximum Jaw
Opening
Altitude
Relative Humidity
Pollution Degree
Safety Compliance
EMC Compliance
Do not use the current clamp if it is damaged. Inspect the case before using the
current clamp. Look for cracks or missing plastic. Pay particular attention to the
insulation surrounding the connectors.
Inspect the clamp jaw before each use. No cracks, missing, loosen or weaken parts
shall be observed. Ensure there is insulation surrounding the jaw.
Inspect the cable without exposing the metal to ensure insulation.
Do not operate the current clamp around explosive gas, vapor, or dust.
Do not exceed the rated voltage/current as marked on the current clamp. Use with
extreme caution when working around bare conductors. Accidental contact with the
exposed conductors could result in electric shock.
Always keep your hand behind the finger guard of the clamp jaw.
When servicing the current clamp, use only specified replacement parts.
Use with caution when working above 30 V ac rms, 42 V peak, or 60 V dc.
These voltages pose a shock hazard.
Do not operate the current clamp if the cover is removed or loosened.
Use individual protective equipment if working in installations with accessible
hazardous live parts.
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired
Do not operate the current clamp around flammable gases or fumes, vapor, or wet
environments.
Do not connect to the BNC output or the banana plug to any power sources.
Use proper terminals, function, and range for your measurements.
Current Clamp
1 A to 400 A ac
0.5 A to 400 A
< 3
10 kHz
294 g
1500 ± 20 mm
32 mm
Up to 2000 m
Max 80% RH for temperature up to 35 °C, decrease linearly to 50% RH at 55 °C
2
Safety Compliance Certified by CSA (Canada & USA) for IEC/EN/UL 61010-1 2nd Edition & EN/
IEC 61010-2-032
Certified to IEC/EN 61326:2002, CISPR 11, and equivalents for Group 1, Class A
Specification
Load impedance
Operating Temperature
Storage Temperature
Measurement Category
Dimensions (HxWxL)
Maximum conductor size
Warm-up time
Warranty
Current Clamp
> 1 MΩ, < 100 pF
–40 °C to 55 °C (–40 °F to 131 °F)
–40 °C to 70 °C (–40 °F to 158 °F)
CAT III 600 V; Pollution degree II
44 mm (H) x 92 mm (W) x 188 mm (L)
30 mm
Immediately upon power on
One year

Advertisement