ZOLL Aed Plus Administrator's Manual page 41

Defibrilator
Hide thumbs Also See for Aed Plus:
Table of Contents

Advertisement

Immunity test
IEC 60601 test
(cont'd)
level (cont'd)
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
The ISM (industrial, scientific and medical) bands between 150 KHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70
MHz.
b
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the
frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable
communications equipment could cause interference if it is inadvertently brought into patient areas.
For this reason, an additional factor of 10/3 is used in calculating the recommended separation
distance for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the ZOLL AED Plus is used exceeds the applicable RF compliance level
above, the ZOLL AED Plus should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating
the ZOLL AED Plus.
d
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
ZOLL AED PLUS Administrator's Guide
Compliance
Electromagnetic environment -
level (cont'd)
guidance (cont'd)
Recommended separation distance
10 V/m
d = 1.20
d = 2.30
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
√P
80 MHz to 800 MHz
√P
800 MHz to 2.5 GHz
b
c
should
d
31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents