Intended Use - Miele PG 8527 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

This Miele washer-disinfector is
designed to clean, rinse, disinfect and
dry re-usable medical devices,
laboratory devices and medical and
laboratory accessories.
Whatever items are being processed,
always comply with the particular
manufacturer's instructions (EN ISO
17664).
Typical applications include:
– surgical instruments,
– endoscopic instruments,
– intensive care and anaesthesiology
equipment,
– baby bottles and nipples,
– supply and disposal containers,
– OR shoes,
or
– laboratory equipment for research
and production,
– laboratory equipment for analytical
and preparation applications,
– microbiology and biotechnology
labware.
The type of laboratory glassware can
include anything from beakers, bootles,
cenrifuge tubes, Erlenmeyer flasks,
funnles, pipettes, Petri dishes and
round bottom flasks.
In this operating manual, the general
term labware/instruments is used to
cover a range of equipment and items.
For standardized results, it is preferable
to machine-wash lab/surgical
instruments and equipment.
When disinfection is required to protect
staff and/or patients, the preference is
for thermal disinfection, e. g., with the
DESIN vario TD process.
– An exception are
temperature-sensitive OR shoes which
can be processed using the
CHEM-DESIN program. –
According to the A
in EN ISO 15883-1, thermal disinfection
occurs at 80°C (+ 5°C, - 0°C) with
10 min. holding time (A
90°C (+ 5°C, - 0°C) with 5 min. holding
time (A
3000), depending on the
0
disinfection result required. The
efficacy standard A
inactivation of HBV.
Regional and/or other official directives
like e. g., CSA Z314.8 Decontamination
of reusable Medical Devices are
applicable.
The cleaning parameters should always
be optimally matched to the type of
soiling and labware/instruments
involved.
The processing chemicals used also
should be matched to the particular
cleaning needs / analysis / analytical
methods involved.

Intended Use

concept described
0
600), or at
0
3000 includes the
0
11