Ophthalmology

Using the E 929/1 mobile unit,
ophthalmological surgical instrument
sets can be cleaned and disinfected
automatically.
For rinsing, use demineralized water
with a conductivity of ~15 ZS/cm
(microsiemens per centimetre).
In addition, the rinse water must have a
low endotoxin and pyrogen content!
The upper level is equipped with
separate connections for hollow
instruments, e.g. rinse and suction
grips and cannulae.
Silicone holders and stoppers fitted on
the mesh tray fix and secure the
instruments to the inlet hose of the
mobile unit.
The lower level of the mobile unit is
fitted with E 441/1 inserts or E 142
mesh trays to prepare instruments
without a hollow body.
,
In washer-disinfectors used to
process small-calibre
ophthalmological instruments,
plastic mesh must not be used.
The mobile units for
ophthalmological surgical instrument
sets come with their own instructions
for use.
,
Mobile unit E 929 (for
ophthalmological surgical
instruments) must only be used in
washer-disinfectors that do NOT
have a dispensing system with a
flow rate of 465 mL/min installed.
,
Ophtalmological instruments
must not be treated in
washers-disinfectors with steam
condenser with heat recovery
(DK+WRG option)!