To Replace the Power Line Fuse

The power line fuse is located within the multimeter's fuse- holder
assembly on the rear panel. The multimeter is shipped from the factory
with a power- line fuse installed (according to country of destination). See
Table 5-
the same size and rating.

Table 5-1 Type of supplied fuse (according to country of destination)

1 Disconnect the power cord. Depress tab 1 and
2 Rotate the line voltage selector and reinstall
U3401A User's and Service Guide
1. If you determine that the fuse is faulty, replace it with one of
Type of fuse(time-lag,low breaking fuse)
0.25A, 250V, 5x20 mm
0.125A, 250V, 5x20 mm
2 and pull the fuse holder from rear panel.
it back, so that the correct voltage appears in
the fuse holder window.
Input line voltage
100 V to 120 V
220 V to 240 V
2 Remove the line voltage selector from fuse
1
2
4 Replace fuse holder assembly in rear
panel.
Disassembly and Repair
Part number
A02-62-25592-3U
A02-62-25648-1U