5-4 Application Options Menu

Updating Signal Generator Patterns

• This procedure requires firmware (Package Version) V1.14 or greater.
• Updating the patterns could take up to an hour to complete.
Note
• For some applications, downloading the patterns makes the unit busy for
Prepare the USB Flash Memory Device
1. The USB flash memory device must be formatted as either FAT or FAT32 file system.
2. The USB flash memory device should have at least 1 GB of free space.
3. Copy the 'tx' folder containing all of the patterns to the top level directory of the USB
device.
Update the S412E Patterns
This process needs to be repeated for each LMR analysis mode where patterns
Note
need to be updated (i.e. P25, P25p2, NXDN, DMR2, TETRA, PTC).
1. Power on the S412E and insert the USB flash memory device into one of the USB ports
on the LMR Master.
2. Enter the desired analysis mode for the patterns being loaded. For example NXDN,
PTC, P25, P25p2, TETRA, or DMR2 mode.
3. Execute the following sequence:
a. Access the System Menu by pressing the Shift key, then press the 8 key on the
keypad.
b. In the System menu, select Application Options > Load Patterns, then press Enter
when ready to start loading patterns.
Figure 5-12. Load Patterns Dialog
c. Wait for the unit to end the loading process (this can take several minutes), press
Enter as directed, and the patterns will be loaded.
d. Press the Esc key when done.
4. Repeat
Step 2
5-10
as long as 15 minutes, but the process does not have to be closely
monitored.
Load Patterns
Press Enter to Start Loading Patterns
and
Step 3
for each analysis mode desired.