3-4 Vector Network Analyzer
4. Press the Limit Alarm key to turn on or off the Limit Alarm.
Figure 3-4.
Single Limit Lines

Set Up Distance Domain

This option is fully described in the Vector Network Analyzer Measurement Guide.
Refer to
Appendix A, "Measurement
on the documentation disc and as free downloads from the Anritsu Web site
(refer to
"Additional Documentation" on page

Calibration Considerations

Various calibrations are available in the LMR Master. Press Shift-2 (Calibration) to open the
calibration menu. Press Start Cal and follow the instruction onscreen. 1-Port calibration is the
simplest and requires three connections during calibration. 2-Port calibration requires
four calibration connections and corrects for the transmit port match.
VNA calibration requires external precision OSL calibration components. 2-Port calibration
also requires a through connector
The Vector Network Analyzer Calibration menu is the same one that is used in the
Vector Voltmeter menu (Option 15). Refer to section "Calibration Menus" in
Chapter 6 of the Vector Network Analyzer Measurement Guide (refer to
Note
Appendix
calibration parameters are shared between the Vector Network Analyzer and
Vector Voltmeter mode, and that some parameters are different because they were
optimized for the specific mode application.
3-6
Guides". Measurement Guides are available
(Figure 3-5 on page
A) for more information about the calibration menu. Note that some
PN: 10580-00318 Rev. P
1-1).
3-7).
Quick Start Guide
S412E UG