Calibrate Ch2 0V; Calibrate Ch2 10V; Calibrate 100Ohm Rtd - Omega FC-21 User Manual

Flow computer
Table of Contents

Advertisement

FC-21 Flow Computer
Sub-menus
8.2.12

Calibrate ch2 0V

Calibrate CH2 0V
Vin=TB1-5
Submenu Group
Calibrate ch2 0V
Calibrate ch2 0V
Calibrate ch2 0V
Iin=TB1-5
Calibrate ch2
8.2.13

Calibrate CH2 10V

Iin=TB1-5
Submenu Group
Calibrate ch2
Calibrate ch2
Calibrate ch2
Iin=TB1-5

Calibrate 100ohm RTD

Calibrate 100ohm RTD
8.2.14
Calibrate 100 ohm
JMP TB1-6,7 100R=7,8
RTD
Submenu Group
Calibrate 100ohm RTD
Calibrate 100ohm RTD
Calibrate 100ohm RTD
JMP TB1-6,7 100R=7,8
Display
GND=TB1-4
ENTER
0 CALIBRATING ——
*** DONE ***
GND=TB1-4
STOP
GND=TB1-4
ENTER
0 CALIBRATING ——
*** DONE ***
GND=TB1-4
STOP
Advance to
ENTER
0 CALIBRATING ——
*** DONE ***
STOP
Advance to
Calibrate 4mA out
To Calibrate: Connect Voltage Source (+)
TB1-5, (-) TB1-4. Input 0V and press Enter.
This message is displayed during calibration.
This message is displayed when the 0V
calibration is finished.
The display will automatically return to the
Calibrate CH2 0V top-level menu. Press the
Down arrow key to advance to the CH2 10V
calibration.
10V
To Calibrate: Connect Voltage Source (+)
TB1-5, (-) TB1-4. Input 10V and press Enter.
10V
This message is displayed during calibration.
10V
This message is displayed when the 10V
calibration is finished.
The display will automatically return to the
10V
Calibrate CH2 10V top-level menu. Press the
Down arrow key to advance to the 100 ohm
RTD calibration.
To Calibrate: Connect a jumper wire between
TB1-6 and TB1-7, Place a 100 ohm 0.1%
resistor between TB1-7 and TB1-8. Press
enter to calibrate.
This message is displayed during calibration.
This message is displayed when the RTD
calibration is finished.
The display will automatically return to the
Calibrate 100 ohm RTD top-level menu.
Press the Down arrow key to advance to the
4mA out calibration.
50
Notes