Liquid Mass

Liquid Mass
Illustration
Calculations
3.3 Liquid Mass
Measurements:
Actual volume is measured by the flow element (DP transmitter,
Flowmeter). Temperature is measured by the temperature transmitter. A
density transmitter can be used for direct density measurements.
Calculations:
• The density and mass flow are calculated using the reference density
and the thermal expansion coefficient of the liquid (see "SET FLUID
PROPERTIES" submenu)
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total, Temperature,
Density
• Analog Output
Rate, Total, Temperature or Density
• Pulse Output
Total
• Relay Outputs
Rate, Total or Temperature Alarms
Applications:
Monitoring mass flow and total of any liquid. Flow alarms are provided via
relays and datalogging is available via analog (4-20mA) and serial outputs.
TOTAL
RATE
PRE 1
TEMP
START
1
2
3
4
STOP
GRAND
SCROLL
PRE 2
DENS
6
7
8
9
Orifice Plate
with DP Transmitter
Volume Flow
As calculated in section 3.1
Mass Flow
Mass Flow = volume flow * density
PRINT
CLEAR
MENU
5
TIME
HELP
ENTER
0
Temp./Dens.
Transmitter
9
FC-21 Flow Computer
TOTAL
RATE
PRE 1
TEMP
PRINT
START
1
2
3
4
5
TIME
STOP
GRAND
SCROLL
PRE 2
DENS
6
7
8
9
0
Flowmeter
Temp./Dens.
Transmitter