SECTION 20—4-WAY VALVE/MANIFOLD ASSEMBLY

Replacing Pilot Valve Poppets and O-Rings

To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: For this procedure, refer to FIGURE 20.2 on page 70.
1. Unplug unit.
2. Remove cabinet. Refer to Cabinet on page 23.
3. Remove the 4‐way valve/manifold assembly. Refer to Replacing 4‐Way Valve and/or 
Manifold Assembly on page 65.
4. Locate pilot valve on 4‐way valve/manifold assembly.
5. Hold pilot valve stem with the flat blade screwdriver and turn the 9/16‐inch locknut 
counterclockwise one complete turn.
6. Unscrew the pilot valve stem assembly from the manifold while leaving the coil with 
yoke and locknut intact on stem.
7. Remove pilot valve poppet from inside the pilot valve stem.
DO NOT remove the coil yoke from the coil. The washer between the bottom of the yoke 
and the bottom of the coil MUST not be removed. Otherwise, components may be 
damaged in reassembly.
8. Set aside the coil with yoke and pilot valve stem assembly with the wires still intact.
DO NOT use sharp tools to remove O‐ring from plastic manifold opening. Damage to the 
plastic manifold and/or plastic manifold airflow passage may occur. Damage to airflow 
may result in improper oxygen delivery and harm to the user. Refer to Detail "A" in 
FIGURE 20.2.
9. Remove poppet and O‐ring from manifold opening (Detail "A").
10. Discard old poppet and O‐ring.
11. Install new O‐ring into the opening in the plastic manifold housing (Detail "A").
DO NOT overtighten pilot valve stem when installing it into plastic manifold. Damage to 
the rubber ʺOʺ ring and/or plastic manifold may occur.