Do you have a question about the Perfecto2 V and is the answer not in the manual?
Questions and answers
charlene deans
March 27, 2025
oxygen level will not increase past 1
1 comments:
Mr. Anderson
March 27, 2025
The provided context does not contain information explaining why the oxygen level on the Invacare Perfecto2 V would not increase past 1.
This answer is automatically generated
charlene deans
March 27, 2025
oxygen level does not increase past 1
1 comments:
Mr. Anderson
March 27, 2025
The Invacare Perfecto2 V oxygen level may not increase past 1 due to several possible reasons:
1. Restricted Airflow – The unit may be placed in a cluttered area or too close to drapes, restricting airflow. Moving it to a well-ventilated area can help. 2. Tubing Issues – Kinked, punctured, or loosely connected tubing can reduce oxygen flow. Checking and replacing tubing or connections as needed may resolve the issue. 3. Power Supply Problems – The unit should not be connected to an extension cord. Ensuring it is plugged directly into a proper power source can help. 4. System Failure – If the system detects an issue and displays a red or yellow warning light, it may indicate low oxygen concentration or another malfunction. Following the manual’s troubleshooting steps or contacting the supplier may be necessary. 5. Service Life Limitation – If the unit has exceeded its expected service life (three years of operation), its performance may degrade, requiring maintenance or replacement.
Checking these factors can help identify and resolve the issue.
Page 1
₂ ₂ ₂ Invacare® Perfecto ™V Oxygen Concentrator IRC5PO2V en HomeFill® System Compatible User M anual This manual M UST be given to the user of t he product. BEFORE using this product, r ead this manual a nd save for future reference.
Page 2
All r ights reserved. Republication, d uplication or modification in whole or in part i s prohibited without prior w ritten permission from Invacare. Trademarks a re identified by ™and ®. All t rademarks a re owned by or licensed to Invacare Corporation or its s ubsidiaries u nless o therwise noted.
Invacare® Perfecto ₂™V Oxygen Concentrator Symbols in Documentation 1 General General W arning Sign 1.1 Symbols The background color i nside the Signal w ords a re used in this m anual a nd apply to hazards o r unsafe triangle is yellow on product l abels.
DO NOT dispose of i n household waste Below 73% 1.2 Intended Use Electrical H azard The intended function and use of t he Invacare® Perfecto ₂™V Oxygen Concentrator is to provide supplemental o xygen to patients Manufacturer with respiratory disorders, b y separating nitrogen from room air, b y way of a molecular sieve.
DANGER! Risk of I njury or Death ₂™V The intended function and use of t he Invacare® Perfecto Depending on their medical c ondition, p atients o n flow Oxygen Concentrator is to provide supplemental o xygen to patients rates g reater t han 5 L/min may be at a n increased risk with respiratory disorders, b y separating nitrogen from room air, b y for serious i njury or death in the event o f f ailure.
Safety 2 Safety 2.1 Label L ocations Serial number l abel is l ocated on the resonator intake assembly. Specification Label i s l ocated on the back of t he concentrator a t t he base. 1193322-A-04...
Invacare® Perfecto ₂™V Oxygen Concentrator 2.2 General G uidelines DANGER! Risk Of D eath, I njury, O r Damage From Fire In order to ensure the safe installation, a ssembly and operation of t he Textiles, oil o r p etroleum substances, grease, greasy concentrator these instructions M UST be followed.
Page 9
– Close supervision is necessary when the nasal c annula accessories. Accessories d esigned by o ther is used by or near children and/or impaired persons. manufacturers h ave not b een tested by Invacare and are not r ecommended for use with Invacare products. WARNING! –...
Page 10
Risk of I njury Risk of I njury or Damage A change in altitude may affect t otal o xygen available to Invacare oxygen concentrators a re specifically designed you. To prevent o xygen deprivation: to minimize routine preventive maintenance. To prevent –...
Safety 2.3 Radio Frequency Interference WARNING! Risk of I njury or Damage To reduce the risk of i njury or p roduct d amage from interference with wireless e quipment: – Keep concentrator a t l east 9 .8 ft ( 3.0 m) a way from wireless c ommunication equipment s uch as w ireless home network devices, mobile phones, cordless phones a nd base stations, w alkie-talkies, e tc.
Invacare® Perfecto ₂™V Oxygen Concentrator 3 Components Oxygen Outlet Flowmeter 3.1 Component Identification Elapsed Time Meter Oxygen P urity I ndicator Lights/Fault and Power Indicator Lights Circuit B reaker Power S witch 1193322-A-04...
Page 13
Components Rear View Power C ord HomeFill O utlet F itting Cabinet F ilter H is t o be used only f or The HomeFill outlet f itting filling oxygen cylinders w ith the HomeFill home oxygen compressor. The outlet f itting does not a ffect c oncentrator performance.
Accessories 4 Accessories 4.1 Optional A ccessories There are many different t ypes of h umidifiers, o xygen tubing, c annulas and masks t hat c an be used with this d evice. You should contact your l ocal h ome care supplier f or r ecommendations o n which of these devices w ill b e best f or you.
Invacare® Perfecto ₂™V Oxygen Concentrator Check for any obvious d amage to the carton or its c ontents. If 5 Setup damage is evident, n otify the carrier, o r your local d ealer. Remove all l oose packing from the carton.
Usage 6 Usage WARNING! Risk of I njury or Damage To avoid injury or d amage from airborne pollutants 6.1 Introduction and/or fumes a nd for optimal p erformance: Your o xygen concentrator i s i ntended for i ndividual u se indoors. –...
Invacare® Perfecto ₂™V Oxygen Concentrator 6.3 Setting Up the Concentrator WARNING! Risk of I njury or Damage Plug in power c ord to an electrical o utlet. Failure to properly install t he humidifier bottle or other Connect H umidifier ( if s o prescribed).
Page 19
Usage Insert a flathead screwdriver in the plate groove on the top edge Humidifier B ottle Humidifier Bottle without Cap of t he filter access d oor C and gently pry the filter access d oor with Cap off. Pull u p and remove the humidifier bottle adapter Remove cap A from bottle B.
Invacare® Perfecto ₂™V Oxygen Concentrator 6.3.2 Turning the Concentrator On Select t he location for the humidifier bottle/adapter assembly: • When placing the humidifier bottle in the compartment on the concentrator, a ttach the humidifier b ottle/adapter tube G to the oxygen outlet c onnector H on the concentrator.
Usage If t he flowrate on the flowmeter ever falls below 0.5 L/min D for more than about o ne minute, t he Potential O bstruction Alert will b e triggered. This i s a rapid beeping of t he audible alarm. Check your tubing or accessories f or blocked or kinked tubing or a defective humidifier bottle.
Invacare® Perfecto ₂™V Oxygen Concentrator 6.3.6 Explanation of O xygen Purity Indicator Lights Indicator Explanation Symbol Light Purity (LED) Immediately GREEN call supplier. light w ith Oxygen sensor i s YELLOW unknown malfunctioning. You light may continue to use flashing the concentrator.
DANGER! Risk of I njury or Damage At a minimum, p reventive maintenance MUST be performed Invacare oxygen concentrators a re specifically designed according to the maintenance record guidelines. In places to minimize routine preventive maintenance. To prevent with high dust o r s oot l evels, maintenance may need to injury or d amage: be performed more often.
Invacare® Perfecto ₂™V Oxygen Concentrator Periodically clean the concentrator’s c abinet a s f ollows: Remove the filter a nd clean as n eeded. Environmental conditions t hat m ay r equire more Use a damp cloth, or s ponge, with a mild detergent s uch as Dawn dish washing soap to gently clean the exterior case.
Maintenance 7.5 Cleaning and Disinfection Between Patients WARNING! Risk of I njury or Damage To prevent i njury f rom infection or d amage to concentrator: – Only qualified personnel should perform cleaning and disinfection of t he oxygen concentrator a nd accessories b etween patients.
Invacare® Perfecto ₂™V Oxygen Concentrator 7.6 Preventive Maintenance Checklist Model N o: Serial N o: ON EACH INSPECTION Record Date of S ervice Record Elapsed Hours o n Hour M eter Clean Cabinet F ilter(s) ( Refer to Cleaning the Cabinet F ilter.) Check Prescribed L/min.
8.2 Wear and Tear 8 After Use Invacare reserves t he right t o ask for a ny item back that h as a n alleged defect i n workmanship. See Warranty that s hipped with the 8.1 Recycling Information product f or s pecific warranty information.
Invacare® Perfecto ₂™V Oxygen Concentrator 9 Troubleshooting 9.1 Troubleshooting SYMPTOM PROBABLE CAUSE SOLUTION Alarm: Main Power Loss: No lights i lluminated. Power cord not p lugged in. Insert p lug into outlet. No power a t o utlet. Inspect h ouse circuit b reakers o r f uses. If p roblem recurs, Short B eeps, l ong pause Tripped circuit b reaker.
Page 29
Troubleshooting SYMPTOM PROBABLE CAUSE SOLUTION Alarm: Potential O bstruction Alert GREEN light i lluminated. Possible internal o bstruction in the Inspect f or k inks o r b lockages. Correct, c lean or r eplace oxygen path. Kinked or b locked item.
5 psi ± 0.5 psi ( 34.5 kPa ± 3.45 kPa) Low Flow Alarm: 0.5 to 5 L/min (maximum). For f lowrates l ess t han 1 L/min, we recommend the use of t he Invacare Pediatric Flowmeter A ccessory (IRCPF16). Potential O bstruction Alert: 0 L/min to 0.5 L/min...
Page 31
Technical data Power C onsumption: 325 W Typical Pressure Relief M echanism Operational a t: 30 psi t o 45 psi ( 207 kPa to 310 kPa ) Change in maximum recommended flow when back pressure of 7 kPa is 0.7 L/min (Back Pressure of 7 kPa applied) applied: Filters:...
Page 32
Invacare® Perfecto ₂™V Oxygen Concentrator Electrical: No extension cords. Placement: No closer than 12 in (30.5 cm ) f rom any wall, f urniture, d raperies o r similar s urfaces t o assure sufficient a ir f low. Avoid deep pile carpets and heaters, r adiators or hot a ir registers.
LIMITATIONS A ND EXCLUSIONS: THE FOREGOING from Invacare or a dealer, w ith a copy of t he seller’s invoice required WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED for c overage under t his w arranty.
Page 34
Invacare® Perfecto ₂™V Oxygen Concentrator THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER, W HETHER EXPRESS OR IMPLIED, I NCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, A ND THE SOLE REMEDY FOR VIOLATIONS OF ANY...
Page 36
Invacare Corporation Manufacturer One Invacare Way Invacare Corporation Elyria, O hio USA 2101 E. L ake Mary Blvd. 44035–2125 Sanford, F L 32773 Tel: 440–329–6000 Tel: 407–321–5630 Tel: 800–832–4707 Technical S ervices Tel: 440-329-6593 Tel: 800-832-4707 www.invacare.com Manufacturer Invacare Rehabilitation Equipment ( Suzhou) C o., L td.
Need help?
Do you have a question about the Perfecto2 V and is the answer not in the manual?
Questions and answers
oxygen level will not increase past 1
The provided context does not contain information explaining why the oxygen level on the Invacare Perfecto2 V would not increase past 1.
This answer is automatically generated
oxygen level does not increase past 1
The Invacare Perfecto2 V oxygen level may not increase past 1 due to several possible reasons:
1. Restricted Airflow – The unit may be placed in a cluttered area or too close to drapes, restricting airflow. Moving it to a well-ventilated area can help.
2. Tubing Issues – Kinked, punctured, or loosely connected tubing can reduce oxygen flow. Checking and replacing tubing or connections as needed may resolve the issue.
3. Power Supply Problems – The unit should not be connected to an extension cord. Ensuring it is plugged directly into a proper power source can help.
4. System Failure – If the system detects an issue and displays a red or yellow warning light, it may indicate low oxygen concentration or another malfunction. Following the manual’s troubleshooting steps or contacting the supplier may be necessary.
5. Service Life Limitation – If the unit has exceeded its expected service life (three years of operation), its performance may degrade, requiring maintenance or replacement.
Checking these factors can help identify and resolve the issue.
This answer is automatically generated