SECTION 14—COOLING FAN
15. Connect spade connectors to fan terminals on side of cooling fan (Detail "B").
16. Move resonator housing back into position (Detail "A").
17. Secure the resonator housing to the soundbox with the three mounting screws. Two in
the top and one on the back near the bottom of the housing (Detail "A").
18. if necessary, reconnect the intake valve on the compressor assembly.
19. Reinstall control panel. Refer to Removing Control Panel on page 53.
20. Reinstall compressor inlet HEPA filter. Refer to Replacing the Compressor Inlet HEPA
Filter on page 26.
21. Plug unit in and check for leaks. Refer to Leak Test on page 76.
22. Run unit for at least 30 minutes. If unit runs properly, replace cover. If not, call
Invacare Technical Service.
23. Reinstall cabinet. Refer to Cabinet on page 23.
™
Perfecto
V
2
56
Part No 1164934