SECTION 4—TROUBLESHOOTING
SYMPTOM
Unit Operating,
Internal Status Indicators:
RED: Three Flashes
GREEN: One Flash
Control Panel Indicators:
RED: Off
YELLOW: Flashing
GREEN: On
Low-Flow Alarm does not
activate on flows less than 0.5
L/min.
Perfecto
V
2
PROBABLE CAUSE
Internal repairs required.
System leak.
Defective check valves.
22
SOLUTION
Replace SensO
circuit board. Refer to
2
Replacing P.C. Board on page 58.
Repair leak in product tank, regulator,
tubing, fittings, or flow meter.
Replace check valves. Refer to Replacing
Check Valves on page 46.
Part No 1164934

Hide quick links: