Summary of Contents for Invacare Platinum POC1-100B
Page 1
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile POC1–100B, POC1–100B-EU, POC1–100B-AZ, POC1–100B-C en Oxygen Concentrator Service Manual DEALER: Keep this manual. The procedures in this manual MUST be performed by a qualified technician.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
All manuals and user guides at all-guides.com Contents 6.4 Cleaning and Disinfection Between Patients ... 43 6.5 Preventative Maintenance Checklist ....44 6.6 Checking O Purity .
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile Refer to the following table for the meaning of symbols 1 General marked on the equipment and/or packaging. 1.1 Symbols General Warning Sign Signal words are used in this manual and apply to hazards...
Page 5
All manuals and user guides at all-guides.com General DO NOT Reuse (Single Patient Use Only) Class II Equipment This symbol refers to the cannula which is for single patient use only. Power On/Off Latex Free Data Output Terminal Indoor Use ONLY This symbol is included on the AC DO NOT dispose of in household waste power adapter.
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile On/Off Button Transport and Storage Humidity Press for one second to turn concentrator on or off. Flow Selection Button This product complies with Directive 93/42/EEC concerning medical devices. Press to adjust the pulsed oxygen flow to your prescribed setting.
All manuals and user guides at all-guides.com General 1.2.2 Input/Output Connections Battery Gauge Indicates the charge status of the battery(s). Refer to Reading the Battery Gauge for Installed and Uninstalled Battery(s) in Setup for more details. External Power Indicator The orange indicator light next to the plug symbol illuminates when an external power source is connected to the concentrator.
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 1.2.3 Filters Air enters the concentrator through an air intake filter located on the carry bag. This filter prevents hair and other large particles in the air from entering the unit. Before you...
All manuals and user guides at all-guides.com Safety 2 Safety DANGER! Risk of Death, Injury Or Damage From Fire Textiles, oil or petroleum substances, grease, 2.1 General Guidelines greasy substances and other combustibles are easily ignited and burn with great intensity in In order to ensure the safe installation, assembly and oxygen enriched air and when in contact with operation of the concentrator these instructions MUST be...
Page 10
Any other accessories have oxygen under pressure. To avoid fire, death, not been tested by Invacare and are not injury or damage: recommended for use with Invacare products. – Avoid creation of any spark near oxygen –...
Page 11
All manuals and user guides at all-guides.com Safety WARNING! WARNING! Risk of Death, Injury or Damage Risk of Injury or Damage To prevent injury or damage from cord misuse: To avoid injury or damage from airborne – DO NOT move or relocate concentrator by pollutants and/or fumes and for optimal pulling on the cord.
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 2.2 Radio Frequency Interference Medical Electrical Equipment needs to be installed and used according to the EMC information in this manual. This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601-1-2.
All manuals and user guides at all-guides.com Setup 3.2 Changing the Date and Time 3 Setup This procedure describes the process to change the 3.1 Viewing the Date and Time date and time. This procedure displays the date and time. The concentrator must be Off and connected to external power to perform this procedure.
Page 14
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 4. Press the flow selection button D to increase the selected value by one until the maximum is reached and then rolls over to the lowest value. 5. Press the display backlight button to decrease the selected value by one until the minimum is reached and 1.
All manuals and user guides at all-guides.com Setup 8. When the time and date have been adjusted as desired, press the power button to highlight Set. 9. Press the flow selection button to save the date and time and to exit the time and date screen. 10.
Page 16
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 3. Press and release the On/Off button A to enter the Language Change Screen. The proposed language selection is highlighted. 4. Press the flow selection button D to move to the next language selection.
Page 17
All manuals and user guides at all-guides.com Setup 6. Press the flow selection button to set the display language to the highlighted language and exit the Language Change Screen. If you do not want to change the display language, press the On/Off power button to highlight “CANCEL”.
4.2 Service and Replacement Parts 6. Remove old patient outlet filter K. 7. Clean all exposed surfaces of both halves of the oxygen Refer to the parts catalog at www.invacare.com for outlet port L with a clean alcohol pad. service and replacement part information.
Only use fuses provided by Invacare. Be careful not to cross the threads when starting to install the cap. The DC power cable does not work unless it is assembled correctly.
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 4.6 Replacing the Battery Pull Ring 1. Use pliers to unscrew and remove the end piece E from the DC power cable F. 1. Lift the rounded end of the existing pull ring.
Page 21
All manuals and user guides at all-guides.com Service 3. Remove the existing pull ring from the battery pack. 5. Slide the new pull ring into the mount. 4. Press the new pull ring down to engage into the mount. 6. Continue sliding the new pull ring into the mount until it is fully engaged.
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 5.1.1 Viewing the Current Alarm Code and Software 5 Troubleshooting Version Number 5.1 Alarm Codes This procedure displays the alarm code of any active alarm condition with the software version number.
All manuals and user guides at all-guides.com Troubleshooting Sample Alarm Code Display Screens Software Version Bootloader Version 5.1.2 Viewing the Alarm Code History The initial “0x” portion of the alarm code is not This procedure displays the 20 most recent alarm shown in the alarm code history screens.
Page 24
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x3201 LPA_noBreathDetected No Breath The display shows: 1. The concentrator 1. Verify cannula is Detected "No Breath Detected...
Page 25
2. If alarm persists, provide between contact Invacare 15 to 30 minutes for service. of additional operation before reaching 0% state of charge. Low Battery...
Page 26
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Battery 0xb20 LPA_batteryTooHot- The display shows: 1. The concentrator 1. Remove the hot Too Hot ToCharge "Cannot Charge Allow...
Page 27
(Running) Warm" power and at least to warm. one of the battery 2. If alarm persists, temperatures is contact Invacare less than or equal for service. to –15° C Unit Hot 0x5203 LPA_unitHot The display shows: 1. The concentrator is 1.
Page 28
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Low Oxygen 0xa20a LPA_lowOxygenPurity The display shows: 1. The concentrator 1. Ensure cannula is Purity "Low Oxygen See...
Page 29
3. Change outlet maximum duration filter. of 120 seconds 4. If alarm persists, before it becomes contact Invacare a prolonged warm for service. up alarm state. a. Unit running without cannula connected -or- b. Inlet airflow blockage -or- c.
Page 30
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x1100 SDA_depletedBattery Depleted The display shows: 1. The concentrator is 1. Supply unit with Battery "Shut Down Battery...
Page 31
2. Change the temperatures is battery. less than or equal 3. If alarm persists, contact Invacare –20° C. for service. Unit Too 0x1103 SDA_unitTooHot The display shows: 1. The concentrator is 1. Allow unit to cool.
Page 32
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name Low Oxygen 0x1105 SDA_lowOxygenPurity The display shows: 1. The concentrator 1. Ensure cannula is Purity (shut "Shut Down Oxygen...
Page 33
2. Disconnect all power source is power sources being connected then reconnect to to the unit. a power source. 3. If alarm persists, contact Invacare for service. 0x110c No Breath SDA_noBreathDetected The display shows: 1. An inhalation has 1. Verfy cannula is Detected "Shut Down No Breath...
Page 34
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x110a Internal ricEven- The display shows: 1. PCB or software 1. If alarm persists, Fault tQueueEmptyAlarm "Shut Down Internal fault.
Page 35
0x1116 fanIsNotMovingAlarm Internal The display shows: 1. Cooling fan or PCB 1. If alarm persists, Fault "Shut Down Internal fault. contact Invacare Fault" for service. 0x1117 fanDriverAttemptI2cBus- Internal The display shows: 1. PCB or software 1. If alarm persists, Fault TimeOutAlarm "Shut Down Internal...
Page 36
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x1120 highPressureAlarm Internal The display shows: 1. Blockage in outlet 1. Remove any kinks Fault "Shut Down Internal gas stream.
Page 37
Name 0x1124 serialeeErrorReadAlarm Internal The display shows: 1. PCB or software 1. If alarm persists, Fault "Shut Down Internal fault. contact Invacare Fault" for service. 0x1128 lcdErrMessageQueue Internal The display shows: 1. PCB or software 1. If alarm persists, Fault "Shut Down Internal...
Page 38
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x1130 errRAMTestFailed Internal The display shows: 1. PCB or software 1. If alarm persists, Fault "Shut Down Internal fault.
Page 39
0x1136 dataflashErrRecevingMes- Internal The display shows: 1. PCB or software 1. If alarm persists, Fault sage "Shut Down Internal fault. contact Invacare Fault" for service. 0x1137 dataflashErrSavingAlarm Internal The display shows: 1. PCB or software 1. If alarm persists, Fault "Shut Down Internal...
Page 40
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile General Alarm Alarm Code Name Display Message Possible Reason Solution Alarm Code Name 0x113e bSPotVVerifyAlarm Internal The display shows: 1. PCB or software 1. If alarm persists, Fault "Shut Down Internal fault.
Page 41
0x1145 adc1SelSample- Internal The display shows: 1. PCB or software 1. If alarm persists, Fault GroupAlarm "Shut Down Internal fault. contact Invacare Fault" for service. 0x1146 adc1StartMeasurementA- Internal The display shows: 1. PCB or software 1. If alarm persists, Fault larm "Shut Down Internal...
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 6.3 Cleaning the Air Intake Filter Screen 6 Maintenance CAUTION! 6.1 Wear and Tear Risk of Damage To avoid damage from clogging: Refer to the User Manual for the list of normal wear and –...
All manuals and user guides at all-guides.com Maintenance 6.4 Cleaning and Disinfection Between Patients 3. Disinfect the surfaces of the concentrator and accessories using Clorox® 4 in One Disinfectant & Cleaning and disinfection must be performed between Sanitizer or substantially equivalent product. patients and as necessary.
Clean Air Intake Filter Screen. Inspect Cabinet of Concentrator and AC Adapter for Damage. Do not use if damage found. Return to Invacare for repair. Inspect Carry Bag Strap(s) and hardware for Damage. Replace strap or carry bag if damage is found.
Page 45
All manuals and user guides at all-guides.com Maintenance EVERY 4,380 HOURS, EVERY 3 YEARS, AND BETWEEN PATIENTS, WHICHEVER COMES FIRST Check Oxygen Purity Perform Functional Test DURING PREVENTATIVE MAINTENANCE SCHEDULE AND BETWEEN PATIENTS Check Intake Filter Screen for damage. Replace carry bag if damage is found. Check/Replace Patient Outlet Filter Inspection periods shown as hours is in reference to hours of concentrator operation since last date of service.
87% or above after a five minute warm-up period. 8. If oxygen purity level is 87% or less, check the air intake filter screen and clean if necessary. If still getting an oxygen purity level less than 87%, contact Invacare for service. 1187375-B...
All manuals and user guides at all-guides.com Maintenance 6.7 Function Test 11. Breathe with the cannula to ensure the unit pulses at each pulse flow setting. Confirm that the pulse 1. Install a battery in the unit and connect to AC power icon is shown on the display screen with each oxygen source using the AC power adapter.
All manuals and user guides at all-guides.com Invacare® Platinum® Mobile 7 After Use 7.1 Disposal Recycle DO NOT dispose of in household waste This product has been supplied from an environmentally aware manufacturer who complies with the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EC. This...
Page 49
All manuals and user guides at all-guides.com Notes...
Page 50
All manuals and user guides at all-guides.com Notes...
Page 51
Ireland • Tel: (353) 1 810 7084 • Fax: (353) 1 810 7085 • ireland@invacare.com • www.invacare.ie Italia Invacare Mecc San s.r.l., Via dei Pini 62, I-36016 Thiene (VI) • Tel: (39) 0445 38 00 59 • Fax: (39) 0445 38 00 34 • italia@invacare.com • www.invacare.it Nederland Invacare AE, Galvanistraat 14–3, NL–6716 BZ Ede •...