Service | Digital Control Unit; Replacing The Front Housing - Invacare Softform Active 2 Rx Service Manual

Powered hybrid support surface
Hide thumbs Also See for Softform Active 2 Rx:
Table of Contents

Advertisement

4 Service | Digital Control Unit

4.1 Replacing the Front Housing

R
efers to spare parts kit SP1633237
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
14 mm Open-ended wrench
1.
Switch off the control unit and disconnect from
mains supply.
2. Turn the unit upside down.
3. Loosen and remove the four screws A.
4. Remove the bottom housing.
5. Loosen and remove the two screws B fixing the
motor to the front housing.
6. Loosen and remove the three screws C fixing
the compressor to the front housing.
60129676-A
A
A
Fig. 4-1
B
Fig. 4-2
C
Fig. 4-3
Invacare
®
Softform Active
®
2 Rx
A
A
B
C
C
7