SECTION 11—REGULATOR
9. Adjust until pressure reads a steady five p.s.i. ± 0.5 p.s.i.
10. Once required pressure is achieved, reinstall cabinet. Refer to Removing Cabinet on
page 23.
11. Refer to Troubleshooting on page 16 if you are unable to adjust or maintain five p.s.i.
± 0.5 p.s.i.
NOTE: After ten minutes of run time, retest output pressure of concentrator to insure proper
operation and/or spec levels.
™