Invacare Perfecto2 V IRC5PO2V Service Manual page 40

Homefill compatible
Table of Contents

Advertisement

SECTION 7—COMPRESSOR
Compressor
To avoid personal injury or property damage, rotate the fan by hand prior to connecting 
the unit to a power source. Check for suction at the air inlet by placing your finger over 
the port as you rotate the fan. You should feel a slight suction with each rotation of the 
fan. If you DO NOT feel a suction but you feel or hear a thump as you turn the fan, DO 
NOT connect the unit to a power source. Review the assembly procedure for possible 
error.
NOTE: For this procedure, refer to FIGURE 7.6 on page 41.
1. Reinstall compressor. Refer to Removing/Replacing Compressor Assembly on 
page 31. 
2. Run unit and check for leaks around compressor hoses and compressor tubes. Refer 
to Leak Test on page 76.
3. Take pressure readings to ensure proper system pressures. Refer to Checking Sieve 
Bed Pressure on page 86.
FIELD SERVICE PARTS LIST FOR GSE 280D2 COMPRESSOR
ITEM
QTY.
NO.
PER ASM
1
2
1
3
1
4
1
5
1
6
7
8
9
2
10
2
11
2
12
1
13
2
14
15
Not Shown
1
Perfecto
V
2
WARNING
QTY.
PER UNIT
2
Connecting Rod, Eccentric and Bearing Assembly
2
2
2
2
1
8
2
4
4
4
2
4
2
2
2
40
DESCRIPTION
Piston Cup Seal
Screw - Piston Cup Retainer
Piston Cup Retainer
Cylinder
Head
Head Screw
Valve Plate Assembly
Valve Flapper Keeper Strip
Square Washer
Valve Flapper - Top & Bottom
Valve Plate
Valve Flapper Screw
Fan - Black
Valve Plate O-Ring - Round
Head Seal O-Ring
Part No 1164934

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents