2
Making Measurements

Alerts and Warning During Measurement

Voltage alert

WA R N I N G
52
For your own safety, please do not ignore the voltage alert. When the
multimeter gives you a voltage alert, immediately remove the test
leads from the source being measured.
This multimeter provides a voltage alert for voltage
measurement in both auto and manual range modes. The
multimeter starts beeping periodically once the measured
voltage exceeds the V-ALERT value set in the Setup mode.
Immediately remove the test leads from the source being
measured.
By default, this feature is turned off. Be sure to set the
alerting voltage according to your requirement.
The multimeter will also display
hazardous voltage when the measured voltage is equal to or
greater than 30 V in all three DC V, AC V and AC+DC V
measurement modes.
For a manually selected measurement range, when the
measured value is outside the range, the display will
indicate OL.
as an early warning for
U1253A User's and Service Guide