Current Terminals Test; Figure 6-2 Current Terminal Error Message - Agilent Technologies U1253A User's And Service Manual

True rms oled multimeter
Hide thumbs Also See for U1253A:
Table of Contents

Advertisement

6
Performance Tests and Calibration

Current terminals test

N O T E
136
This test determines whether the input warning for the
current terminals is functioning properly.
Turn the rotary switch to any non- off position other than
. Insert the tests leads to the A and COM terminals.
An error message Error ON A INPUT (as shown in
will be displayed on the secondary display, and a continuous
beep will persist until the positive lead is removed from the
A terminal.
Before conducting this test, make sure the beep function is not disabled in
Setup.

Figure 6-2 Current terminal error message