Selecting Display With The Hz Button; Table 1-9 Selecting Display With The Hz Button - Agilent Technologies U1253A User's And Service Manual

True rms oled multimeter
Hide thumbs Also See for U1253A:
Table of Contents

Advertisement

1
Getting Started Tutorial

Selecting display with the Hz button

Table 1-9 Selecting display with the Hz button

Rotary switch position (Function)
(For
, press
voltage)
(Default is DC voltage)
(Press
twice to select AC+DC voltage)
(Default is DC voltage)
22
The frequency measurement function is able to detect the
presence of harmonic currents in neutral conductors and
determine whether these neutral currents are the result of
unbalanced phases or non- linear loads.
• Press
for current or voltage measurements — voltage or current
on the secondary display, and frequency on the primary
display.
• Alternatively, pulse width (ms) or duty cycle (%) can be
displayed on the primary display by pressing
This allows simultaneous monitoring of real- time voltage
or current with frequency, duty cycle, or pulse width.
• Hold
current reading on the primary display.
Primary display
Frequency (Hz)
Pulse width (ms)
to select AC
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
Frequency (Hz)
Pulse width (ms)
Duty cycle (%)
to enter the Frequency measurement mode
for more than 1 second to resume voltage or
U1253A User's and Service Guide
again.
Secondary display
AC V
DC V
AC+DC V
DC mV

Advertisement

Table of Contents
loading