Performance Tests and Calibration
6

Other recommendations for calibration

Specifications are only guaranteed within the specified
period from the last calibration. Agilent recommends that
complete readjustment should always be performed at
whatever calibration interval you select. This will ensure that
the U1253A true RMS OLED multimeter remains within its
specifications until the next calibration. This calibration
criterion provides the best long- term stability.
During performance verification tests, only the performance
data is collected; these tests do not guarantee that the
instrument will remain within the specified limits. The tests
are only for identifying which functions need adjustment.
Please refer to
"Calibration count"
on page 167 and verify
that all adjustments have been performed.