5.
Cut the three cable ties shown in Figure 58. Disconnect the two braided hoses from the pressure and
vacuum ports of the ATF. Cap the three ATF module ports using tight fitting vinyl caps or vinyl electrical
tape.
Once the braided hoses are disconnected from the ATF Module the
CAUTION
ATF Module is exposed to the atmosphere. Cap the ATF Module ports
immediately in order to minimize exposure to debris and humidity
which may result in ATF Module damage.
Figure 58: Removal of Compressor Pressure and Vacuum Hoses from ATF Module.
Personal Ambulatory Oxygen System Provider Technical Manual
Cable
Tie
To ATF Module
Vacuum Port
To ATF Module
Pressure Port