REMOVE AND REPLACE HEPA FILTER

Replace the HEPA filter annually, or more often as needed. To replace the HEPA filter, follow these steps.
WARNING:
NOTE:

Internal Filters

HEPA Filter replacement:
1. Cut cable tie and Disconnect the tubing at the top of the filter carriage.
Personal Ambulatory Oxygen System Provider Technical Manual
DO NOT use any petroleum based or other lubricants. A spontaneous
and violent ignition may occur if oil, grease or other petroleum
substances come into contact with oxygen under pressure. Keep
these substances away from the oxygen system, tubing and
connections and any other oxygen source.
Always cut the heads of cable ties to avoid damaging the tubing.
Figure 19: Tubing on Filter Carriage.
44
PN 20631679 Rev D