CAIRE Eclipse 5 Provider Technical Maunal page 10

Personal ambulatory oxygen system
Hide thumbs Also See for Eclipse 5:
Table of Contents

Advertisement

The Eclipse 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse 5
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic
±6 kV contact
environment –
guidance
±8 kV air
IEC 61000-4-2
Electrical fast tran-
±2 kV for power supply
sient/burst
lines
IEC 61000-4-4
±1 kV for input/output lines
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0,5 cycle
on power supply
input lines IEC
40 % U
61000-4-11
(60 % dip in U
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A / m
(50/60 Hz) mag-
netic field IEC
61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Personal Ambulatory Oxygen System Provider Technical Manual
Guidance and manufacturer's declaration–electromagnetic immunity
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
N/A
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
T
for 5 sec
3 A / m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30 %.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
T
)
cal commercial or hospital environment. If
T
the user of the Eclipse 5 requires continued
operation during power mains interruptions,
it is recommended that the Eclipse 5 be pow-
T
)
ered from an uninterruptible power supply or
T
a battery.
T
)
T
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
10
PN 20631679 Rev D