The Eclipse 5 Power Cartridge operation time may be affected by several factors such as bolus size, breathing rates, ambient
temperature, age of power cartridge and use over time. The table below describes the typical operating time for a new Power
Cartridge.
If the Eclipse 5 is used in Pulse Dose Mode, there will be longer operating time. The patient should consult their physician for
a prescription for Pulse Dose Mode.
Continuous
Flow Setting
0.5 LPM
1.0 LPM
2.0 LPM
3.0 LPM
NOTE: Battery times will decrease with higher bolus size, breath rate, ambient temperature, Battery age and use over time.
The Power Cartridge packaged with the Eclipse 5 is not fully charged. Before using the Eclipse 5 Oxygen System for the first
time, the Power Cartridge must be fully charged.
Store the power cartridge in a cool, dry location. Do not leave the Eclipse 5 or Power Cartridge in a vehicle or trunk during a
hot or cold day.
When checking the Eclipse 5 as baggage on a commercial airline flight for international travel, remove the Power Cartridge
and properly package.
When shipping the Eclipse 5 for any reason, remove the Power Cartridge from the Eclipse 5.
If the Power Cartridge gets too warm, charging will not begin until the Power Cartridge sufficiently cools. Consider removing
the Power Cartridge to allow for faster cooling.
The Power Cartridge operating time is longer if the Eclipse 5 is operated in Pulse Dose Mode. (Refer to Battery Consumption
Chart above.)
The typical time to recharge the Power Cartridge to achieve 80% capacity from a fully discharged Power Cartridge is 1.8 hours
to 5.0 hours, dependent upon the flow setting.
The capacity of the Eclipse 5 Power Cartridge is determined by electronics and the Eclipse 5 software.
Personal Ambulatory Oxygen System Provider Technical Manual
TYPICAL NEw PowER CARTRIDGE oPERATING TIMES
Battery Duration
4.4 hours
3.7 hours
2.0 hours
1.3 hours
Battery Consumption Chart
Pulse
Battery Duration
Setting
(12 BPM)
-
-
16mL 1.0
5.4 hours
32mL 2.0
5.1 hours
48mL 3.0
4.9 hours
64mL 4.0
4.0 hours
80mL 5.0
3.7 hours
96mL 6.0
3.5 hours
128mL 7
2.5 hours
160mL 8
2.0 hours
192mL 9
1.7 hours
(At nominal
temperature
of 25ºC/77°F)