CAIRE Eclipse 5 Provider Technical Maunal page 11

Personal ambulatory oxygen system
Hide thumbs Also See for Eclipse 5:
Table of Contents

Advertisement

The Eclipse 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse 5
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
a
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Eclipse 5 is used exceeds the applicable RF compliance level above, the
Eclipse 5 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Eclipse 5.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
Personal Ambulatory Oxygen System Provider Technical Manual
Guidance and manufacturer's declaration–electromagnetic immunity
Compliance level
3 Vrms
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Eclipse 5, including cables, than
the recommended separation distance cal-
culated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P
80 MHz to 800 MHz
d = 2.3 P
800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site sur-
vey,
should be less than the compliance level
a
in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following sym-
bol:
11
PN 20631679 Rev D