Pulse Dose Mode Specifications
Pulse Settings
Trigger Sensitivity
Adjustable Bolus Rise Time
Trigger Criteria
Minimum time between
breaths
Response to Missing Breaths
Pulse Dose
Setting
1.0
2.0
3.0
4.0
5.0
6.0
7
8
9
NOTE: Bolus volume decreases as breath rate exceeds published range.
Power Accessory Specifications
AC Power Supply
Input Voltage
Input Power
Output Voltage
Output Power
Power Cartridge (Battery)
Output voltage
Capacity
Nominal Power Cartridge Life
Power Cartridge Recharge Time
Personal Ambulatory Oxygen System Provider Technical Manual
1.0 to 6.0, in 8mL increments up to 96 mL; 7-9 in 32mL increments up to 192 mL
Adjustable between settings of 1 (most sensitive) to 3 (least sensitive)
Adjustable settings of Fast(factory setting), Medium, or Slow
•
Cannula pressure has dropped below the trigger point (typically between 0.135—0.37
cm of H
O of negative pressure)
2
•
At least 1¼ seconds has passed since the last pulse began
1.25 seconds (max. 3 consecutive breaths)
While in Pulse Dose Mode, the Eclipse 5 is always monitoring for breath detection.
After 15 seconds of no breath detected, the Eclipse 5 "delivers" Continuous Flow at the
last Continuous Flow setting. After another 15 seconds, the Eclipse 5 stops delivering
Continuous Flow and waits 15 seconds for a breath. The Eclipse 5 will stay in this
modality until a breath is detected. (See Power Cartridge Conservation Feature, page 26.)
Bolus Size
Power Cartridge (Battery)
(± 15%) mL
16
32
48
64
80
96
128
160
192
100-240VAC, 50-60 Hz
245-260 VA
28 VDC
200W
14.8 VDC
Quantity (2) 97.5 W-hrs batteries
(Each containing 7.92 grams equivalent Lithium content)
80% Capacity after 500 Charge/Discharge cycles
1.8 to 5.0 hours, dependent on flow setting, to achieve 80% capacity
from a fully discharged Power Cartridge
AC Power Supply and
Max Breath
Rate
40
40
40
40
37
31
23
18
15
DC Power Supply
Input Voltage
Output Voltage
Output Power
DC Power Supply
Max Breath
Rate
40
40
40
31
25
20
15
12
10
11.5-16VDC
26 VDC
150W Max
PN 20631679 Rev D
7