6. Disconnect the cable from the Control Panel as follows:
a.
Grasp the circuit board firmly between your fore fingers and thumb.
b.
Grasp the head of the Membrane Panel Overlay cable in your other hand.
c.
Firmly pull the cable away from the board. Pull the cable straight out. Do not rock the
connector back and forth. This may damage the pins of the header.
CAUTION
7. Disconnect the silicone tube from the oxygen outlet tube at the top of the cover. First cut the cable tie,
then pull the silicone tube off the oxygen outlet tube.
NOTE:
CAUTION
Personal Ambulatory Oxygen System Provider Technical Manual
Do not disconnect the Membrane Panel Overlay cable by pulling on
the ribbon cable.
Figure 14: Removing the Membrane Panel Overlay Ribbon Cable.
Always cut the heads of cable ties to avoid damaging the tubing.
Avoid possible eye injury by wearing protective eyewear or shielding
the eyes from possible flying debris.
Membrane Panel
Overlay Cable
Connection
PN 20631679 Rev D
41