12. Remove caps from ATF Module.
13. Connect the silicone tube from the Product Tank into the product port of the ATF and secure joints with
cable ties as shown on Figure 31.
14. Use cable tie to secure the silicone tube to the braid tube to ensure that there is NO kinking on the
bend as shown on Figure 32. Hand-tighten the cable tie. Inspect the tubing to ensure the tubing is not
kinked.
15. Perform product test prior to installing cover and again after cover replacement.
16. Installation of the Unit Cover on the unit is described in the Remove and Replace of the Unit Cover.
Personal Ambulatory Oxygen System Provider Technical Manual
66
PN 20631679 Rev D