MALFuNCTION CODES

If a malfunction occurs in the Eclipse 5, the device will stop, the Red LED on the front panel will light and the buzzer will sound for 10
seconds and then silence. The LCD will display one of the following Malfunction Codes:
Malfunction
Invalid RESET
I/O Port Failure
RAM Failure
FLASH Failure
EEPROM Failure
IPC Watchdog Timeout
Compressor Motor Too Hot
PCB Too Hot
Compressor Motor Stalled
Battery Communication Failure
IPC Failure
Product Pressure Sensor Failure
Breath Pressure Sensor Failure
Ultrasonic Failure
Product Temperature Sensor Failure
Loss of Product Tank Pressure
Purity Calibration Data Failure
Flow Calibration Data Failure
Breath Sensitivity Data Failure
Hour of Operation Data Failure
Invalid Assembly Option
Ambient Pressure Sensor Failure
Unidentified Failure
Step 1: Remove all power sources; battery and external power. Wait 20 seconds. Reconnect power to the Eclipse 5. Turn the
device to the ON position. If resolved, continue use.
Step 2: If unresolved, call Chart Technical Support for assistance.
Personal Ambulatory Oxygen System Provider Technical Manual
FAIL code
FAIL 80
Check for secure communication cables, Verify functionality of PM and CS PCBs
FAIL 81
Check for secure communication cables, Verify functionality of PM and CS PCBs
FAIL 82
Check for secure communication cables, Verify functionality of PM and CS PCBs
FAIL 83
Check for secure communication cables, Verify functionality of PM and CS PCBs
FAIL 84
Check for secure communication cables, Verify functionality of PM and CS PCBs
FAIL 90
Check for secure communication cables, Verify functionality of PM and CS PCBs
Allow Eclipse 5 to Cool for 30 Mintues. Check compressor fan functionality. Check to ensure vents
FAIL 91
are not obstructed. Verify functionality of the Compressor Assy
Allow Eclipse 5 to cool for 30 Minutes. Verify functionality of PM fan. Check to ensure vents are
FAIL 92
not obstructed
Check Compressor wires for damage. Verify functionality of PM PCB. Verify functionality of the
FAIL 94
Compressor Assy
Verify functionality of the Battery Bridge PCB. Check Battery and Battery Bridge PCB for pin dam-
FAIL 95
age. Verify functionality of battery. Verify functionality of PM pcb
FAIL A0
Verify functionality of PM PCB
FAIL A1
Verify functionality of CS PCB
FAIL A2
Verify functionality of CS PCB
FAIL A3
Verify functionality of CS PCB. Check for water intrusion from humidifier
FAIL A4
Verify functionality of DB9 cable harness. Verify functionality of CS PCB
FAIL A5
Check for leaking, cracked HEPA filter or damaged tubing. Verify ATF motor is turning
FAIL A6
Check for leaking, cracked HEPA filter or damaged tubing. Verify functionality of CS PCB
FAIL A7
Verify functionality of Proportional Valve. Verify functionality of CS PCB
FAIL A8
Check for damaged tubing. Verify functionality of CS PCB
FAIL A9
Verify functionality of CS PCB
FAIL AA
Remove Power to Reset
FAIL AB
Verify functionality of CS PCB
FAIL FF
Remove Power to Reset
Recommended Action
60
PN 20631679 Rev D