CAIRE Eclipse 5 Provider Technical Maunal page 64

Personal ambulatory oxygen system
Hide thumbs Also See for Eclipse 5:
Table of Contents

Advertisement

5.
Cut the cable ties to the braided tubes that connect into the ATF pressure and vacuum ports as shown
in Figure 32. Disconnect the braided tubes.
5a. Install port caps.
6.
Unscrew the 3 screws (M4x16 Pan Head Machine Screw, P/N 6961-416-SEQ) and remove the screws and
washers (M4 Flat washer, P/N 6985-04-SEQ) as shown below.
7.
Lift the ATF. Take off the 3 grommets. Disconnect the wire harness from the gearbox of the ATF.
8.
Remove the ATF.
9.
Install 3 grommets into the mounting slots on the ATF as shown below. Position the ATF into the
bottom cover but DO NOT install the mounting screws yet. Route the 16" Silicone Tube under the ATF as
shown on Figure 35 on next page.
Caps
Personal Ambulatory Oxygen System Provider Technical Manual
Remove
Screws
Figure 33: Remove ATF Module from Eclipse 5 Case.
Grommet
Figure 34: ATF Module Grommet Installation
Caps
64
PN 20631679 Rev D