Nominal
Range at
Full Scale
(V)
24 Hrs
±10
0.0437
Nominal
Range at
Full Scale
(V)
24 Hrs
±10
0.0512
© National Instruments Corporation
Table A-5. NI 6115 Analog Output DC Accuracy Information
Absolute Accuracy
% of Reading
90 Days
1 Year
0.0445
0.0454
Table A-6. NI 6120 Analog Output DC Accuracy Information
Absolute Accuracy
% of Reading
90 Days
1 Year
0.0520
0.0529
Voltage Output
Ranges .................................................... ±10 V
Output coupling...................................... DC
Output impedance .................................. 50 Ω ±5%
Current drive .......................................... ±5 mA
Output stability....................................... Any passive load
Protection .............................................. Short-circuit to ground
Power-on output voltage
(before software loads calibration values)
NI 6115 ........................................... ±400 mV
NI 6120 ........................................... ±80 mV
Initial power-up glitch
Magnitude ....................................... ±2 V
Duration .......................................... 200 ms
Temp
Absolute
Offset
Drift
Acc. at Full
(mV)
(%/°C)
Scale (mV)
8.9
0.0006
Temp
Absolute
Offset
Drift
Acc. at Full
(mV)
(%/°C)
Scale (mV)
1.9
0.0006
A-13
Appendix A
Specifications
Relative Accuracy
Theoretical Resolution
(mV)
13.5
4.88
Relative Accuracy
Theoretical Resolution
(µV)
6.7
305.2
NI 6115/6120 User Manual