Counter N Internal Output And Counter N Tc Signals; Routing Counter N Internal Output To An Output Terminal; Frequency Output Signal; Routing Frequency Output To A Terminal - National Instruments NI USB-621x User Manual

Bus-powered m series usb devices
Hide thumbs Also See for NI USB-621x:
Table of Contents

Advertisement

Counter n Internal Output and Counter n TC Signals

Frequency Output Signal

Default Counter/Timer Pinouts

Table 9-5. Default NI-DAQmx Counter/Timer Pins for USB-6210/6211/6215 Devices
Counter/Timer Signal
CTR 0 SRC
CTR 0 GATE
CTR 0 AUX
CTR 0 OUT
© National Instruments Corporation
Some of these options may not be available in some driver software.
Counter n TC is an internal signal that asserts when the counter value is 0.
The Counter n Internal Output signal changes in response to Counter n TC.
The two software-selectable output options are pulse on TC and toggle
output polarity on TC. The output polarity is software-selectable for both
options.
Routing Counter n Internal Output to an Output
Terminal
You can route Counter n Internal Output to any PFI <4..7> or PFI <12..15>
terminal.
The Frequency Output (FREQ OUT) signal is the output of the frequency
output generator.

Routing Frequency Output to a Terminal

You can route Frequency Output to any PFI <4..7> or PFI <12..15>
terminal.
By default, NI-DAQmx routes the counter/timer inputs and outputs to the
PFI pins, shown in Tables 9-5 and 9-6.
CTR 0 A
CTR 0 Z
Default Terminal Number (Name)
9-29
Chapter 9
1 (PFI 0)
2 (PFI 1)
1 (PFI 0)
6 (PFI 4)
1 (PFI 0)
3 (PFI 2)
NI USB-621x User Manual
Counters

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ni usb-621 series

Table of Contents