Chapter 11
Isolation and Digital Isolators
NI USB-621x User Manual
Add an AC return path from the device isolated ground to earth ground. For
non-isolated systems, an AC return path is only needed for high or source
impedances. An AC return path can be created by connecting a capacitor
between the device's isolated ground and earth ground.
11-4
ni.com