C3 Mode - Span Adjustment - KERN GAB 3K1DM Service Manual

Hide thumbs Also See for GAB 3K1DM:
Table of Contents

Advertisement

3) Operation
1.
Stating Test
2.
Scale No. and Decimal point indication mode
- Press the [ TARE ] key → 1
- Example: ASIA 6kg (5g/2g) Multi interval X1=2, X2=4
2.1 Scale No. Setting
- Press the [ MODE ] key one time ( X1=2)
2.2 Decimal point indication Setting
- Press the [ ZERO ] key → 2
- Press the [ MODE ] key three times ( X2=4)
- When the C3 mode is required, press the TARE key.
- When finishing, press the
record the data to the memory, and enter the Ending
Test Mode.
2.1.8 C3 Mode – Span Adjustment
1) Operation
1.
Stating Test
2.
Span Adjustment mode
2.1 Press the TARE key two times.
2.2 Press the MODE key → displays original A/D data
(The normal range for original A/D data is 1000 to
25000 counts).
2.3 Press the MODE key without load on the weigh platter
→ Approx. 5000 count is diplayed.
Operation
st
digit flashes
nd
digit flashes
Memory Switch to
Operation
- 12 - GAB/GAS-SH-e-0711
Display
X2 X1
X2 X1
X2 X1
X2 X1
Display

Advertisement

Table of Contents
loading

Table of Contents