8845A/8846A
Programmers Manual
STATus:PRESet
Description Sets all bits to zero in the Meter's Questionable Data enable register.
Examples
*CLS
Description Sets all bits to zero in the Meter's status byte register and all event
Examples
*ESE <enable value>
Description Enables or disables bits in the Meter's Standard Event enable register.
Parameters <n>
Examples
*ESE?
Description Retrieves the value in the Meter's Standard Event enable register.
Examples
*ESR?
Description Retrieves the value in the Meter's Standard Event register.
Examples
*OPC
Description Sets bit 0 (Operation complete) in the Meter's Standard Event register.
Examples
*OPC?
Description Returns a one to the Meter's output buffer.
Examples
*PSC <n>
Description Enables or disables the clearing of the Status Byte Enable register and
Parameters <n>
62
STAT:PRES
registers. Also clears the error queue.
*CLS
= 0 to 255
*ESE 160
*ESE?
*ESR?
*OPC
*OPC?
Standard Event Enable register when Meter power is turned on. This
setting is stored in non-volatile memory.
= 0
1
Sets all bits to zero.
Sets all bits to zero and clear error queue.
Binary-weighted decimal value representing the
bits to set in the register. Although this register
has 8 bits and will accept a number from 0 to
255, only bits 0, 2 through 5, and 7 are used.
Sets bits 5 and 7 to one and all other bits to zero.
Retrieves a binary-weighted decimal value
representing the bits set in the Standard Event
enable register.
Retrieves a binary-weighted decimal value
representing the bits set in the Standard Event
register.
Sets bit 0 in the Standard Event register.
A "1" is placed in the output buffer of the Meter.
Disables clearing Status Byte and Standard
Event Enable register at power-up.
Enables clearing Status Byte and Standard Event
Enable register at power-up.