Fluke 8842A Emulation Limitations

Introduction

Due to differences in their design, the 8845A and 8846A (collectively the 884XA) cannot
emulate all the functions and features of the Fluke 8842A. Some functions are compatible
but are not implemented the same in the 884XA meters.
This appendix covers the differences with the 8842A, and gives a list of commands that
are not compatible, and information on those functions that the 884XA will respond
differently to.

Device Dependent Command Set

Table 39 lists Fluke 8842A commands that are not supported in the 8842A emulation
mode and the IEEE-488 response received with they are used.
Table 39. 8842A Device Dependent Commands Not Supported
Command
W6
Enable EOI Only
W7
Disable All Output Terminators
G2
Get Calibration Input Prompt
P2
Put Variable Calibration Value
P3
Put User-Defined Message
C0 – C3
C3 Calibration Commands
Table 40. Fluke 8845A/8846A Commands the Fluke 8842A Doesn't Support
Command
L1
L2
L3

Calibration Commands

The 8845A/8846A does not support instrument calibration in 8842A emulation mode. If
you wish to calibrate the instrument, you must be in 884XA command mode. Any Cn
calibration command will generate an Error 51.
Description
Switches to 884XA command language
Switches to Fluke 45 command language
Switches to Fluke 8842A command language
Appendix D
8842A Emulation Response
Command is ignored
Command is ignored
Error 51 returned
Error 51 returned
Error 51 returned
Error 51 returned
Description
113