Error Code
Error Text
701
CAL Secure Code Case Invalid characters in calibration secure code. The secure code
702
CAL Secure
703
CAL Secure code
704
CAL Secure code too
long
706
CAL Invalid step
number
707
CAL Reference
708
CAL Not calibrating
709
CAL Abort
710
CAL Sequence
711
CAL No measurement
712
CAL Overload
713
CAL Too far
714
CAL Math
715
CAL Timeout
Table 35. Calibration Errors
must contain only characters 0-9 and A-Z. No lower case or
other special characters are allowed.
Cal secured.
Invalid secure code.
Secure code too long.
Invalid calibration step number.
Cal reference value out of tolerance.
CAL? only works if you are calibrating.
Calibration aborted.
Calibration steps out-of-sequence.
No measurements taken during calibration.
Overload at input during calibration.
Calibration measurements out-of-tolerance.
Math error during calibration.
Timeout error during calibration.
Description
Remote Operation
Error Codes
101