Issuing a MEASure? command without specifying a function, puts the
Meter into volts dc and triggers a measurement. You must specify a range
to use a resolution parameter.
Example
MEAS:volt:ac? 10
MEAS:diod? ON
Notes
[1] For ac measurements, resolution is fixed at 6½ digits. The resolution
[2] For frequency measurements, the multimeter uses one "range" for all
[3] For period measurements, the multimeter uses one "range" for all
[4] The range and resolution are fixed for temperature measurements.
[5] The range and resolution are fixed for diode test: 10 Vdc range and 5½
[6] The range and resolution are fixed for continuity tests: 1 kΩ range and
[7] DC measurement voltage divided by dc reference voltage. For dc ratio,
For best results in RATIO, the two input commons must be shorted at the
Meter terminals. The analog filter (AFLTR) should be off.

Using the CONFigure Command

The CONFigure command allows more configuration control than the MEASure
command. As with the MEASure command, the CONFigure command presets the
Meter's parameters for the desired configuration (See Table 18). However, a
measurement does not start automatically, and therefore affords the opportunity to change
measurement parameters prior to triggering a measurement.
The CONFigure command is very useful when most of the parameters preset by
CONFigure are what you want and you only need to change one or two parameters. The
Note
= Returns a measurement after configuring the
Meter for volts ac and setting the range to the 10
volt range.
= Sets diode function using low current and returns
the diode measurement. 9.90000000E+37 is the
value returned for an open diode.
parameter only affects the front panel display.
inputs between 3 Hz, and 300 kHz. With no input signal applied,
frequency measurements return zero.
inputs between 0.33 seconds, and 3.3 µsec. With no input signal
applied, period measurements return zero.
However, placing PT100_385 or PT100_392 into the range field will
set the RTD type. Placing CUST1 into the range field will cause the
Meter to use the values set into the R
digits. When set to ON, the low current parameter sets the diode test
current to 0.1 mA. When OFF, the current is 1 mA. When set to ON,
the high voltage parameter sets the diode test voltage to 10 volts. When
OFF, the voltage is 5 volts. To specify a diode test voltage, you must
also include ON or OFF for the low current parameter.
5½ digits.
connect the reference HI/LO to the HI/LO sense terminals of the Meter
and the measurement voltage to the HI/LO input terminals of the
Meter. Note the specified measurement range applies to the input
terminals only.
Note
Supported SCPI Commands
and Alpha parameters.
0
Remote Operation
39