8845A/8846A
Programmers Manual
SENSe:FUNCtion command provides a means of changing a function's parameters.
Other commands, such as INPut, SENSe, CALCulate, and TRIGger allow for more
detailed control of the Meter's parameters.
The CONFigure command does not initiate a measurement and will need to be followed
by READ? command, or the INITiate and FETCh? commands.
CONFigure[:SCALar][:<function>] [<range |MIN|MAX|DEF> or <RTD type> or <low
current>][,<resolution |MIN|MAX|DEF> or <high voltage>]]
Description Sets the Meter function with the preset parameters (See Table 18) and
The dc portion of the function keyword is optional, as is the VOLTage
portion. For instance, CONF:AC will be interpreted by the Meter as
meaning Volts ac function.
Parameters <function> =
You must specify a range whenever specifying a resolution parameter.
40
configure the Meter using the specified range and resolution. This
command must be followed by a READ? command, or the INITiate and
FETCh? commands to cause the Meter to take a measurement. For the
temperature function, the range variable is replaced by the RTD type
variable.
CAPacitance
CURRent:AC
CURRent:DC
VOLTage:AC
VOLTage:DC
VOLTage:DC:RATio DC voltage ratio
RESistance
FRESistance
FREquency
PERiod
TEMPerature:RTD
TEMPerature:FRTD
DIODe
CONTinuity
<range>
=
a number between the upper and lower limits of the
function.
MIN
MAX
DEF
<low current> = ON or 1
OFF or 0
<low voltage> = ON or 1
OFF or 0
<resolution> =
a number in the same units, not in number of digits, as
the measurement function,.
MIN
MAX
DEF
Note
Capacitance
AC current
DC current
AC voltage
DC voltage
2-wire resistance
4-wire resistance
Frequency
Period
2-wire temperature
4-wire temperature
Diode
Continuity
Lowest range of the function
Highest range of the function
Autorange
Sets diode current to 0.1 mA
Sets diode current to 1 mA.
Sets diode voltage to 5 volts
Sets diode voltage to 10 volts.
Smallest value acceptable
Largest value acceptable
5½ digits
Note
[1]
[1]
[7]
[2]
[3]
[4]
[4]
[5]
[6]

Hide quick links: