8845A/8846A
Programmers Manual
Error Code
-126
-127
-130
-137
-150
-154
-157
-191
-192
-193
-200
-211
-213
-214
-222
-223
-227
96
Table 33. Execution Error Codes and Descriptions (cont.)
Error Text
Numeric real
Channel dimension
invalid
Parameter suffix
Invalid header suffix
Invalid string data
String size
Unmatched brackets
Unknown parameter
Too many dimensions
No entry
Generic execution
Trigger ignored
Init ignored
Trigger deadlock
Illegal data value
Too much data
Invalid function
Description
A floating point (real) numeric parameter value was entered.
Example: SAMP:COUN -13.6
One or more entries in a channel list parameter has an invalid
number of dimensions.
A parameter within the Input Command had the wrong type of
units for the Command Specification. Example
VOLT:DC:RANGE 1A
Too many numeric suffices in the input for the specified
command. As an example, valid arguments for the FETCH?
Command are FETCH?, FETCH1?, FETCH2?, and FETCH3?.
If FETCH4? is entered, error 137 will be generated.
The parameters in the Input Command contained an
unmatched single or double quote. Example: DISP:TEXT 'hello
world".
The length of the entered string exceeds 12 characters.
Example CAL:SEC:CODE "this is an example of a long string".
The parameters in the Input Command contain an unmatched
bracket.
One or more of the parameters contains an unspecified error.
Too many dimensions in entry.
There was no entry to get - the index was beyond the end of
the entries.
General Execution error has occurred.
A Group Execute Trigger (GET) or *TRG was received but the
trigger was ignored. Make sure the multimeter is in the wait-for-
trigger state before issuing a trigger, and make sure the correct
trigger source is selected.
An INITiate command was received but could not be executed
because a measurement was already in progress. Send a
device clear to halt a measurement in progress and place the
multimeter in the idle state.
A trigger deadlock occurs when the trigger source is BUS and
a READ? command is received.
A character string was received, but could not be executed
because the string length was more than 12 characters. This
error can be generated by the CALibration:STRing and
DISPlay:TEXT commands.
Decibels modifier was applied to an selected function that does
not support decibels.

Hide quick links:

Advertisement