If you want the Model 2303 or 2303B to power up with the output on, you must use SAV5,
SAV6, SAV7, SAV8, or SAV9 as the power-on setup. For the SAV5 power-on setup, the power
supply will power up to the SAV0 settings, and the output will be on or off depending on the
output state when the setup was saved as SAV0. For example, assume the output is on and the
setup is saved as SAV0. With SAV0 as the power-on setup, the power supply will power up with
the output off. With SAV5 as the power-on setup, the power supply will power up with the output
on.
Power-On Setups:
NOTE
Models 2303, 2303B and 2303PJ:
SAV0 (output off)
SAV1 (output off)
SAV2 (output off)
SAV3 (output off)
SAV4 (output off)
Models 2303 and 2303B:
SAV5 (SAV0 setup with output on or off)
SAV6 (SAV1 setup with output on or off)
SAV7 (SAV2 setup with output on or off)
SAV8 (SAV3 setup with output on or off)
SAV9 (SAV4 setup with output on or off)
NOTE

Menu

Many aspects of operation are configured from the menu that is summarized in Table 1-2. Use
the following rules to navigate through the menu structure.
NOTE
SAV5 through SAV9 are not available for the Model 2303-PJ.
For GPIB operation, the setups are saved and recalled using the *SAV, *RCL, and
*RST commands. (See Section 8 for details.) The power-on setup is selected using the
SYSTem:POSetup command (Section 10).
The menu key is used to access the menu structure. However, if in remote for IEEE-
488 bus operation ("R" displayed below "ON/OFF") the menu key returns the instru-
ment to LOCAL operation.