11-6
SCPI Tables
Table 11-5

SOURce command summary

Command
[:SOURce]
:VOLTage
[:LEVel]
[:IMMediate]
[:AMPLitude] <n>
[:AMPLitude]?
:CURRent
[:LIMit]
[:VALue] <NRf>
[:VALue]?
:TYPE <name>
:TYPE?
:STATe?
*LIMRELAY (LIMIT RELAY) and TRIPRELAY are available in Models 2303/2303B with firmware revision A06 and higher.
Description
Path to set output voltage:
Specify voltage amplitude in volts: 0 to 15 (1mV resolution).
Query voltage amplitude.
Path to configure current:
Path to configure current limit:
Specify current limit value in amps: 0 to 5 (100µA resolution).
Query current limit value.
Select current limit type (LIMit, TRIP,
LIMRELAY|LIMITRELAY, or TRIPRELAY).*
Query current limit type.
Query state of current limit: 1 = in current limit (for
LIMit type) or output tripped (for TRIP type);
0= not in LIMit/TRIP.
Default
parameter SCPI
0.0
0.25
LIM