● Horizontal stage movement
Problem
Rotating the stage
knobs does not move
the stage.
Pressing the operation
button does not change
the sensitivity
(coarse/fine/extra fine)
of horizontal stage
movement.
XYZ speed interlock is
ON, but the sensitivity of
stage XYZ movement
does not switch
corresponding to the
objective.
Motorized quadrocular tilting tube/Motorized DSC zooming port for quadrocular tube
Problem
The optical path does
not switch.
Motorized zoom does
not operate.
Chapter 2
Cause
The motorized XY stage is not connected
properly.
The motorized XY stage is at the specimen
removal position.
The motorized XY stage is at the specimen
removal position.
Information for objectives is not
configured.
(If the objective information is not
configured, the sensitivity of stage XYZ
movement is not changed.)
Cause
The motorized quadrocular tilting tube is
attached incorrectly.
The quadrocular tilting tube is fastened.
The motorized DSC zooming port is
attached incorrectly.
Troubleshooting
Connect the stage properly.
(→See Chapter 1 "3 Assembly Method - 17 Attach
the stage" in the "Assembly/Maintenance".)
Press the operation button to release the
specimen removal position.
(→See Chapter 5 "1.18.3 Retracting the Elevating
Section and Moving the Stage to the Specimen
Removal Position" in the DS-L3 instruction manual
"Microscope Operation".)
Press the operation button to release the
specimen removal position.
(→See Chapter 5 "1.18.3 Retracting the Elevating
Section and Moving the Stage to the Specimen
Removal Position" in the DS-L3 instruction manual
"Microscope Operation".)
Configure the objective information.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (1) Setting Optical
Elements Installation Information" in the
"Operation".)
Attach it correctly.
(→See Chapter 1 "3 Assembly Method - 12 Attach
the quadrocular tilting tube and DSC zooming port
for quadrocular tube" in the
"Assembly/Maintenance".)
Unfasten it. (Remove the tightening screw for
the quadrocular tilting tube.)
(→See Chapter 1 "3 Assembly Method - 12 Attach
the quadrocular tilting tube and DSC zooming port
for quadrocular tube." in the
"Assembly/Maintenance".)
Attach it correctly.
(→See Chapter 1 "3 Assembly Method - 12 Attach
the quadrocular tilting tube and DSC zooming port
for quadrocular tube" in the
"Assembly/Maintenance".)