Motorized ND filter
Problem
Motorized nosepiece
does not move at all.
The motorized ND filter
does not interlock when
the objective is
switched.
Brightness is not
adjusted appropriately
when the objective is
switched.
Objectives are
configured correctly, but
the brightness is not
adjusted appropriately
when the objective is
switched.
Ergo controller and motorized XY stage
● Button
Problem
The operation button
does not respond. Both
LEDs on the FL-OBJ
toggle switch are off.
The operation button
does not respond.
However, pressing the
FL-OBJ toggle switch
turns its LED on and off.
Chapter 2
Cause
The motorized ND filter cable is not
connected.
The interlock setting is OFF.
Interlocking conditions are not satisfied.
Incorrect configuration of the objectives
The ND filter status set for each objective
is not appropriate.
Focusing and centering of the condenser
is not correct.
Aperture diaphragm is not 75%.
Cause
The ergo controller is not connected
properly.
No function is assigned to the operation
button.
Input from the operation button is disabled.
Corresponding motorized device is not
connected properly.
Troubleshooting
Connect the cable properly.
(→See Chapter 1 "3 Assembly Method - 25 Connect
the motorized unit cable" in the
"Assembly/Maintenance".)
Turn the setting ON.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Check motorized device combination and
setting.
(→See Chapter 3 "19.2 Interlocking Function" in the
"Operation".)
Configure correctly.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (1) Setting Optical
Elements Installation Information" in the
"Operation".)
Set interlock to "Standard". Or, set another
appropriate status.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Configure correctly.
(→See Chapter 3 "7 Focusing and Centering the
Condenser" in the "Operation".)
Set to 75%.
(→See Chapter 3 "19.2 Interlocking Function" in the
"Operation".)
Connect the ergo controller properly.
(→See Chapter 1 "3 Assembly Method - 25 Connect
the motorized unit cable" in the
"Assembly/Maintenance".)
Check and configure the assignment of
function to the operation button.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (3) Configuring the
Button Functions" in the "Operation".)
Check and configure the enable/disable
setting for the operation button input.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (3) Configuring the
Button Functions" in the "Operation".)
Connect the motorized device properly.
49