Problem
The motorized universal
condenser's aperture
diaphragm does not
interlock when the
objective is switched.
The optical module
corresponding to an
objective is not moved
into the optical path
when the objective is
switched.
The aperture diaphragm
is not adjusted
appropriately when the
objective is switched.
Objectives are
configured correctly, but
the aperture diaphragm
is not adjusted
appropriately when the
objective is switched.
Configuration of address
1 is set to "OPEN" and
cannot be changed.
Cannot configure to
"OPEN".
Field diaphragm
Problem
Field diaphragm does
not interlock when the
objective is switched.
The field diaphragm is
not adjusted
appropriately when the
objective is switched.
Objectives are
configured correctly, but
the field diaphragm is
not adjusted
appropriately when the
objective is switched.
Chapter 2
Cause
The interlock setting is OFF.
Interlocking conditions are not satisfied.
Incorrect configuration of the objectives
and modules
The turret was moved manually while the
power is ON.
Incorrect configuration of the objectives
The aperture diaphragm correction value
for objective change is not appropriate.
Focusing and centering of the condenser
is not correct.
Structurally, a filter cannot be attached to
address 1. Therefore, it is preset to
"OPEN" and cannot be changed.
Configuration "OPEN" is exclusive to
motorized universal condenser address 1.
Other addresses cannot be configured to
"OPEN".
Cause
The interlock setting is OFF.
Interlocking conditions are not satisfied.
Incorrect configuration of the objectives
The field diaphragm correction value for
objective changes is not appropriate.
Focusing and centering of the condenser
is not correct.
Troubleshooting
Turn the setting ON.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Check motorized device combination and
setting.
(→See Chapter 3 "19.2 Interlocking Function" in the
"Operation".)
Correct the setting for each.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (1) Setting Optical
Elements Installation Information" in the
"Operation".)
Turn on the power supply again.
Do not move the turret by hand while the
power is ON.
Configure correctly.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (1) Setting Optical
Elements Installation Information" in the
"Operation".)
Set interlock setting to "75%". Or, set another
appropriate value.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Configure correctly.
(See Chapter 3 "7 Focusing and Centering the
Condenser" in the "Operation".)
Use it as it is.
Use it as it is.
Turn the setting ON.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Check motorized device combination and
setting.
(→See Chapter 3 "19.2 Interlocking Function" in the
"Operation".)
Configure correctly.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (1) Setting Optical
Elements Installation Information" in the
"Operation".)
Set interlock to "100%". Or, set another
appropriate value.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Configure correctly.
(See Chapter 3 "7 Focusing and Centering the
Condenser" in the "Operation".)
48
Measure
Measure