Nikon Eclipse Ni-E Instruction Manual page 55

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

Problem
XYZ speed interlock is
ON, but the sensitivity of
vertical stage movement
does not switch
corresponding to the
objective.
Vertical movement is
displayed as
"+13000.000um to
-2000.000um".
Motorized nosepiece
Problem
Motorized nosepiece
does not move at all.
Objectives cannot be
switched smoothly.
The motorized nosepiece
skips addresses.
The stage does not retract
when it should when the
objective is being switched
(or retracts when it should
not).
Motorized universal condenser
Problem
The motorized universal
condenser's turret and
aperture diaphragm do
not move at all.
The center of the optical
module is not aligned
with the optical axis.
The motorized universal
condenser's turret does
not interlock when the
objective is switched.
Chapter 2
Cause
Information for objectives is not
configured.
(If the objective information is not
configured, the sensitivity of vertical stage
movement is not changed.)
The stage vertical movement part is not
attached correctly.
Cause
The motorized nosepiece is attached
incorrectly.
A function to restrict the nosepiece rotation
stop by vertical movement position is set.
Objectives are mounted on rather one side
of the nosepiece.
The rotation stop function from address 1 to
6 (7) is set.
The toggle function is ON.
Incorrect escape setting
Cause
The motorized universal condenser is
attached incorrectly.
Control error. Or, the turret was moved by
hand while the power is ON.
The interlock setting is OFF.
Troubleshooting
Configure the objective information.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (1) Setting Optical
Elements Installation Information" in the
"Operation".)
Attach it correctly.
(→See Chapter 1 "3 Assembly Method - 16 Attach
the substage" in the "Assembly/Maintenance".)
Attach it correctly.
(→See Chapter 1 "3 Assembly Method - 19 Attach
the nosepiece" in the "Assembly/Maintenance".)
Turn off rotation stop or rotate after changing
the vertical movement position.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Attach objective evenly.
Turn off rotation stop.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Turn the toggle function OFF.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Configure correctly.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Attach it correctly.
(→See Chapter 1 "3 Assembly Method - 18 Attach
the condenser" in the "Assembly/Maintenance".)
Turn on the power supply again.
Do not move the turret by hand while the
power is ON.
Turn the setting ON.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
47
Measure
Measure
Measure