■ Controls/buzzer on the main body
Problem
Buttons on the main
body do not respond.
Motorized devices are
not interlocked by
pressing the FUNCTION
button.
Brightness cannot be
adjusted with the
dia-illumination
brightness control knob.
Buzzer does not sound.
When the power switch
is turned on, a series of
short beeps sound, and
the Escape button lamp
blinks.
Motorized devices are
not interlocked by
pressing the MODE
switching button.
Stage vertical movement
Problem
Focus cannot be
adjusted with the focus
knob.
Parfocal correction is
not performed when the
objective is changed.
Focus is not corrected or
is corrected improperly
when parfocal correction
is performed after the
objective is changed.
Pressing the Escape
button does not retract
the stage elevating
section.
Chapter 2
Cause
Operation is disabled.
The MODE function is not set.
Operation is disabled.
The buzzer is set to mute.
The vertical movement counter is
abnormal.
The interlock setting is OFF.
Cause
The focal position is outside the range
limited by the software.
The focus knob is disabled.
The stage elevating section is in the
escape state.
Parfocal correction is OFF.
The focal points of the objectives are not
set properly.
The retracted position is lower than the
lower limit of the vertical motion range.
Troubleshooting
Change the setting.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (3) Configuring the
Button Functions - (3-1) Ni-E microscope button" in
the "Operation".)
Set the MODE function.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (5) Set Mode Function"
in the "Operation".)
Change the setting.
(→See Chapter 3 "23 Operation on DS-L3" -
23.1 Setting Up the Microscope - (3) Configuring
the Button Functions - (3-1) Ni-E microscope
button".)
Change the setting.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (6) Configuring Other
Functions" in the "Operation".)
Press the Escape button and initialize the
vertical movement. When initialization is
complete, the buzzer emits a long buzz.(The
buzzer emits beeps for 20 seconds. If the
escape button is not pressed during this time,
vertical movement will automatically be
initialized.)
Turn the setting ON.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Disable the software limit.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (6) Configuring Other
Functions" in the "Operation".)
Enable the focus knob.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (6) Configuring Other
Functions" in the "Operation".)
Press the Escape button to reset the escape
state.
Turn the setting ON.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Set the correct focal point for each objective.
(→See Chapter 3 "23 Operation on DS-L3" - "23.1
Setting Up the Microscope - (4) Configuring the
Movement" in the "Operation".)
Not escaped because the elevating section is
already away from the retracted position. No
action is necessary.
46