[Fuse replacement method]
WARNING
• To prevent electric shock or malfunctions, always turn off the power of the motorized XY stage controller
(press to the "O" position) and unplug the power cord from the wall outlet before replacing the fuse.
• Use only the specified fuse. Use of other fuse may result in malfunction or fire.
Specified fuse: 250V, 6.3A, time lag φ5.2 x 20 mm (Manufacturer: SOC Corp., Model: ET)
CAUTION
• Check that the contact is not damaged before attaching. If it is damaged, power may not turn on or
overheating may occur.
• Firmly place the fuse in the fuse holder. If it is not firmly placed, overheating or smoke may occur due to
dropped fuse or bad contact.
• After replacing the fuse, make sure the fuse holder is securely attached.
(1)
Push the power switch at the front of the motorized
XY stage controller to "O" position and pull out the
power cord if it is connected.
(2)
Pull out the fuse holder using a flathead screwdriver.
Chapter 1
Assembly
Power off the motorized XY stage controller
22
MADE IN JAPAN
TI-S-CON
100-240V~
2A 50/60Hz
5 6 1 0 0 1
NO.
This device complies with Part 15 of the FCC Rules.Operation is subject to the
following two conditions:
(1) this device may not cause harmful interference,and
(2)
this device must accept any interference received, including interference
that may cause undesired operation.
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.
P O W E
AC IN
R
~
100-240V
2A 50/60Hz
Power switch
AC inlet
with fuse
holder
This device complies with Part 15 of the FCC Rules.Operation is subject
following two conditions:
(1) this device may not cause harmful interference,and
(2) this device must accept any interference received, including interfere
that may cause undesired operation.
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-00
AC IN
~
100-240V
2A 50/60Hz
Fuse
AC250V, T6.3AL
Pulling out the fuse holder